8-76
8.5.3 Preparations Before Using the Valve Diagnosis Tool
Before identifying valve or sensor faults by using the valve diagnosis tool, make preparations as
follows:
1
Connect the airway according to the type sensor or valve to be checked.
Before using the diagnosis test menu, connect the tubes of the anaesthesia machine with
constant flow to check the flow sensor and inspiratory valve. For details, see section
7.3.2 Flow calibration (Service).
2. Ensure that the gas supply pressure complies with the regulations.
3. When the system is in standby mode, choose
Main
>
Service
>
Diagnostic Tests
to
access the diagnostic test menu.
8.5.4 Diagnosing Zero Point Exceptions of Sensors
You can easily determine whether the zero points of all pressure sensors and flow sensors comply
with the regulations.
To diagnose the zero points of the sensors, perform the following operations:
1. Disconnect all gas sources and ensure that the actual values of the sensors are 0.
2. Query the A/D sampling value of each sensor in the menu of the valve diagnosis tool.
This value is the zero point of the sensor.
3. If the zero point of any sensor exceeds the specified range, the VCM is faulty. You need
to replace the VCM.
You can also check the zero points of sensors according to section
6.6.4 Check the Sensor Zero
Point
.
Note
For details about the normal range of zero points of sensors, see section 3.7.3.
8.5.5 Diagnosing Flow Data Exceptions of Sensors
The flow sensor has two sampling tubes. Connection exceptions include:
The two sampling lines are connected reversely.
One sampling line is not connected.
Neither sampling line is connected.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...