8-24
Message
Alarm
Priority
Cause
Solution
CO2 module.
EtCO2
Overrange
Low
The monitored value is
out of the measurable
range.
1. Re-calibrate the CO2 module.
2. Replace the CO2 module.
FiCO2
Overrange
Low
The monitored value is
out of the measurable
range.
1. Re-calibrate the CO2 module.
2. Replace the CO2 module.
CO2 Zero
Failed
Low
The CO2 module is
faulty.
1. Re-zero the CO2 module.
2. Replace the CO2 module.
CO2 Init Error
High
CO2 initialization error
Replace the CO2 module.
Incompatible
CO2 Software
Version
High
The CO2 module is not
compatible with the
software version.
Replace the CO2 module (the M02A
module is not supported).
8.2.9 Mainstream CO
2
module:
Message
Alarm
Priority
Cause
Solution
EtCO2
Overrange
Low
The monitored value is
out of the measurable
range.
1. Re-calibrate the CO2 module.
2. Replace the CO2 module.
FiCO2
Overrange
Low
The monitored value is
out of the measurable
range.
1. Re-calibrate the CO2 module.
2. Replace the CO2 module.
CO
2
Zero
Failed
Low
The CO2 module is
faulty.
1. Re-zero the CO2 module.
2. Replace the CO2 module.
CO2 Init Error
High
CO2 initialization error
Replace the CO2 module.
CO2 Sensor
Error
Low
The sensor of the
mainstream CO2 module
is faulty.
Replace the CO2 module.
CO2 No
Sensor
Low
The sensor of the
mainstream CO2 module
is not connected, or a
communication error
occurs.
Replace the CO2 module.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...