8-59
6. Push the pushrod of the syringe to make the reading of the anaesthesia machine
calibration apparatus rise to 70-90 cmH2O. Stop pushing the pushrod and remain the
relative position between the pushrod and the syringe body unchanged. If the pressure
reading of the anaesthesia machine calibration apparatus decreases no more than 5
cmH2O within 15s, the test item is passed.
II. Leakage test on the pressure sampling line of the flow sensor inside the system (If test item I is
not passed, conduct this test)
1. Fix the test fixture of the circuit transit adapter to the circuit transit block assembly.
2. Connect the pressure sensor interface (positive pressure end) of the anaesthesia machine
calibration apparatus, syringe (pull out the pushrod to a certain position before
installation), and the test fixture of the circuit transit adapter to the corresponding
connectors of the Silicon hose (6x300) (the connector except for connects 1 to 6 on the
test fixture) using the Y-shape tube.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...