2-17
Pin No.
Signal name
Description
10)
NTC1
Lithium battery 1, connecting to thermistor signal inside
11)
GND ground
12)
P15V
Heating wire drive voltage output (Range:12.42V to
15.18V)
13)
HRT22
Pin 1 of heating wire thermistor
14)
HRT21
Pin 2 of heating wire thermistor
15)
GND ground
16)
3.3V
3.3V supply valtage output (Range: 3.135V to 3.465V)
17)
GND ground
18)
GND ground
19)
GND ground
20)
12V2
Second 12V supply valtage output (Range: 0V to 0.4V)
21)
GND ground
22)
GND ground
23)
12V1
First 12V supply valtage output (Range: 10.8V to 13.2V)
24)
15.2VB
15.2V supply valtage output (Range:14.44V to 15.96V)
25)
15.2VB
15.2V supply valtage output (Range:14.44V to 15.96V)
26)
PLAM
Audio buzzer drive signal which drives audio buzzer
directly
27)
SWITCH
Circuit switch indicating whether the circuit is in place
28)
P-FAN+
Drive of heat-sink fan of power board
29)
LED-BAT
Drive output of battery status indicator light
30)
LED-AC
Drive output of AC status indicator light
31)
PCON
Power on/off signal, LVTTL signal. When this signal is
high level, the system is on; when this signal is low level,
the system is off.
32)
BAT2+
Lithium battery 2 input, connecting to the anode of the
battery
33)
BC2
With/without Lithium battery 2 signal. High level
indicates it is with battery; low level indicates it is without
battery.
34)
BAT1+
Lithium battery 1 input, connecting to the anode of the
battery
35)
BC1
With/without Lithium battery 1 signal. High level
indicates it is with battery; low level indicates it is without
battery.
36)
GND ground
37)
P15V
Heating wire drive voltage output (Range:12.42V to
15.18V)
38)
HRT12
Pin 1 of heating wire thermistor
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...