8-27
2. Set the machine following the instructions on the screen, and then select Continue to
perform the automatic circuit leak and compliance test. Progress of the automatic circuit
leak and compliance test is displayed on the screen. As shown in the figure below, select
Cancel to cancel the ongoing test.
3. After the test is completed, confirm whether the automatic circuit leak and compliance
test is passed following the instructions on the screen. If not, deal with it following
"Help" on the screen and the following "Note".
NOTE
If the leak test is not passed, check all possible sources of leaks, including bellows,
respiratory system tubes and CO2 absorber canister. Check whether the sources of
leaks are properly connected and their connectors are damaged.
8.3.2 Manual Circuit Leak Test
This test checks if the pneumatic circuit has leakage in manual ventilation mode. The test items
include APL valve, check valve, CO2 absorber canister, patient tube, flow sensor, and flow sensor
connector.
The respiratory system leak test under manual ventilation mode shall be performed following the
steps below:
1. If the system is being powered on, the system automatically initiates a self-test and
enters the Manual Circuit Leak and Compliance Test screen. Or, from the main screen,
select General > Test Leak/Compliance. Select Manual Circuit Leak Test to enter the
Manual Circuit Leak and Compliance Test screen as follows.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...