8-67
5. Check the canister assembly.
Required tools:
One VT PLUS
One Lucer adapter connector
One syringe
Two
Φ
6 silicon tubes
One PU tube (6X300)
Three breathing tubes
One Y-shape tube
One Y-shape breathing tube
One breathing tube adapter connector
One T-shape hexagon screw wrench
Test procedures:
(1) Disable the system switch.
(2) Remove the Pre-pak assembly and remove the breathing circuit.
(3) Install the Pre-pak assembly.
(4) Remove the sealing parts on two connectors of the canister assembly. Connect two
corrugated pipes and one Y-shape connector to two connectors of the lifting device.
Connect the other end of the Y-shape connector to the breathing tube adapter connector
using another corrugated pipe. Connect the syringe connector, pressure sensor connector
(positive pressure end) of the anaesthesia machine calibration device, and the breathing
tube adapter connector to the Y-shape tube, as shown in the following figure.
(5) Push the pushrod of the syringe to make the pressure reading on the anaesthesia machine
calibration apparatus rise to 30-35 cmH2O. Then, stop pushing. Remain the relative
position between the pushrod and the syringe body unchanged. If the pressure reading of
the anaesthesia machine calibration apparatus decreases more than 10 cmH2O within
30s, the canister assembly encounters a leakage issue. The test is not passed. This step is
required when the bypass of the canister assembly is enabled or disabled.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...