3-22
3.2.7 Monitoring Products Mounting and Electrical Connection
(optional)
1. Mount the monitor (if available) according to the manufacturer’s monitor assembly
instructions.
NOTE
Use of other monitor and mounting hardware is performed by the installer.
2. After mounting a monitor to this device, connect it to one of the AC outlets located on
the rear of this device.
3. Turn on each monitor one at a time. Verify that the circuit breaker holds without
tripping.
4. Dress each line cord neatly along the side of the anesthesia machine or tucked inside the
monitor arm. An optional cable routing kit is available. The cable routing kit contains
three (3) clips, screws, and two (2) Ethernet cables. The clips attach to three (3) sets of
holes on the rear door of this device. Ethernet and power cables can be routed through
the clips.
3.2.8 Anesthesia Module Installation and Waste Gas
Scavenging
1. The respiratory gas monitoring product has an exhaust hole for emitting waste gas. The
exhaust hole on the gas monitor must be connected to the sample gas return port at the
back of this device.
2. Make sure that a tight connecting fitting is attached to the exhaust port of gas monitor.
Make sure that the other end of the same tube is connected to the back of this device.
3. Dress the exhaust tube neatly along the side of the anesthesia machine or tuck it inside
the monitor arm so that it cannot be easily pulled and it does not extend far from the
main chassis.
3.2.9 Negative Pressure Suction
For the installation of negative pressure suction, see 12.8 Negative Pressure Suction Device of
WATO EX-55/65 Operator’s Manual.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...