8-17
Message
Alarm
Priority
Cause
Solution
normally with the
keyboard for 10s
consecutively.
keyboard and CPU board.
3. When necessary, replace the keyboard.
4. When necessary, replace the CPU board.
8.2.6 External AG Module Runtime Alarms
Message
Alarm
Priority
Cause
Solution
AG Hardware
Error
Medium
The AG hardware is
faulty.
Replace the AG module.
O2 Sensor
Error
Medium
The O
2
sensor is
faulty.
Replace the AG module.
External AG
Self Test Error
Low
AG self test error
1. Reconnect the AG module.
2. If the problem persists or the alarm is
repeatedly generated, replace the AG module.
AG Hardware
Malfunction
High
The AG module is not
properly installed or is
faulty.
Replace the AG module.
AG Init Error
High
An error occurs during
AG initialization.
1. Reconnect the AG module.
2. If the problem persists or the alarm is
repeatedly generated, replace the AG module.
AG No
Watertrap
Low
The watertrap falls off
from the AG module.
1. Check the watertrap of the AG module.
2. Replace the watertrap of the AG module.
3. Replace the AG module.
AG Watertrap
Type Wrong
Low
The patient type is
neonate or infant, but
the watertrap type is
adult.
Replace the watertrap with the neonate-type
watertrap.
AG Change
Watertrap
Medium
The AG watertrap
needs to be changed.
1. Check the watertrap of the AG module.
2. Replace the watertrap of the AG module.
3. Replace the AG module.
AG Comm
Stop
High
The AG module is
faulty or encounters a
communication
failure.
1. Replace the communication cable of the AG
module.
2. Replace the AG module.
AG Airway
Occluded
High
The actual pump rate
of the AG module is
lower than 20 ml/min
for more than 1s.
1. Check the sampling of the AG module.
2. Replace the watertrap of the AG module.
3. Replace the AG module.
AG Zero
Failed
Low
Zeroing of the AG
module fails.
1. Zero the AG module again.
2. Replace the AG module.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...