7-55
Calibrate as follows:
1. Make sure that the CO2 module is already warmed up.
2. Select
the
[
Maintenance
] shortcut key
→
[
Factory Maintenance >>
]
→
enter the
required password
→
[
Module Cal. >>
]
→
[
Gas Module Cal. >>
]
→
[
CO2 Module
Cal.
].
3. Check the airway and make sure that there are no occlusions or leaks.
Vent the sampling line to the air and check that the current rate is approximately
150 mL/min.If the deviation is great, it means that the airway is occluded. Check
the airway for occlusions.
Block the gas inlet of the sampling line. The current rate should drop rapidly and
the message of airway occlusion should be prompted. Otherwise, it means that the
airway leaks. Check the airway for leakage.
4. Wait for the sensor temperature to reach and stay at 35ºC.
5. Select
[
Zero
] to start zeroing.
6. Connect the gas cylinder to the sampling line using a T-shape connector, as shown
below.
7. Vent the sampling line to CO2 opening the cylinder pressure relief valve.
8. In
the
[
CO2 Module Cal.
] menu, enter the vented CO2 concentration in the [
CO2
]
field.
9. In
the
[
CO2 Module Cal.
] menu, the measured CO2 concentration, barometric pressure,
sensor temperature and current pump rate are displayed. After the measured CO2
concentration becomes stable, select [
CO2 % Cal
.] to calibrate the CO2 module.
10. After a successful calibration, the screen shows [
Calibration Completed!
]. Otherwise,
the message [
Calibration Failure! Please try again.
] is displayed. In this case, you
need to do the calibration again
.
7.3.8.3 Commonly-encountered Problems and Recommended Actions
Failure description
Possible cause
Recommended action
Calibration is not
completed.
1. The module is damaged.
2. The difference between the set
AG calibration concentration and
the selected standard AG
concentration is too great.
1. Return the module to factory for
repair.
2. The difference between the standard
gas concentration and the set
calibration concentration cannot
exceed 40% of the standard gas
concentration.
AG module
Gas cylinder
Open to the air
Relief valve
Tubing
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...