4-40
4. When
the
system
is standby, select
Main
->
Service
->
Diagnostic Tests
->
Valves
to
access the following menu.
5. Set Safety Valve to On.
6.
Set
PEEP Valve Pressure
to 30 cmH2O.
7. Set
Insp Valve Flow
to the following values: (3±0.5)L/min, (10±1)L/min, (20±1)L/min,
(30±2)L/min, (60±3)L/min
Make sure that the deviation between the measured data of the inspiratory flow sensor, expiratory
flow sensor and ventilator flow sensor and that of the anesthesia machine calibration device must
not exceed 1 L/min or 5% of the measured value of the calibration device, whichever is greater.
Otherwise, perform flow calibration again.
Reinstall the bellows and the water collection cup. When testing the flow accuracy of (60
±
3)L/min, only compare the accuracy of the inspiratory flow sensor and expiratory flow sensor.
It is no need to compare the ventilator flow sensor.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...