7-57
10. If the calibration is finished successfully, the message [
Calibration Completed!
] is
displayed. Otherwise, the message [
Calibration Failure! Please try again.
] is
displayed. In this case, you need to do the calibration again
.
NOTE
If the calibration fails, you can select [
Defaults
] to restore the factory default
calibration values. If the deviation is great, select [Calibrate] again to do a
calibration.
If the calibration still fails, replace the AG module.
7.3.9.3 Commonly-encountered Problems and Recommended Actions
Failure description
Possible cause
Recommended action
Calibration menu is
inaccessible.
The AG module is not fully warmed up
or not in measure mode.
Wait for the AG module to be
fully warmed up and then access
the calibration menu.
Calibration is not
completed.
1. The module is damaged.
2. The difference between the set AG
calibration concentration and the
selected standard AG concentration is
too great.
1. Return the module to factory
for repair.
2. The difference between the
standard gas concentration and
the set calibration concentration
cannot exceed 25% of the
standard gas concentration.
7.3.10 O2 Module Calibration (factory)
7.3.10.1 Preparations
Calibrate the O2 module once a year or when the measured value has a great deviation. Prepare
the following before doing the calibration:
Gas cylinder with 100% O2 and the O2 concentration not less than 99%.
T-shape connector
Tubing
7.3.10.2 Calibration Procedures
NOTE
Do not calibrate the O2 module when there are significant leaks in the airway.
In case of calibration failure, select [
Calibrate
] again to do a calibration.
If the calibration still fails, replace the O2 module.
Calibrate the O2 module, if it has been transported for long distance or if you
suspect it does not work properly.
Follow this procedure to perform a calibration:
1. Make sure that the system is not Standby and the AG module is in [
Measure
] state.
2. Wait for the AG module to be fully warmed up.
3. Select
the
[
Maintenance
] shortcut key
→
[
Factory Maintenance >>
]
→
enter the
required password
→
[
Factory Cal. >>
]
→
[
Gas Module Cal. >>
]
→
[O2
Module
Cal.
].
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...