8-20
Message
Alarm
Priority
Cause
Solution
BIS R Over
Range
Low
The monitored value of
BIS R exceeds the
measurable range.
Replace the BIS module.
BIS High
Imped.
Low
The sensor is not properly
attached to the skin of the
patient, resulting in a high
impedance.
1. Check whether the BIS sensor is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Sensor Off Low
The sensor is detached
from the patient.
1. Check whether the BIS sensor is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
1 High Imped.
Low
The impedance of BIS
electrode 1 is too high.
1. Check whether BIS electrode 1 is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
1 Lead Off
Low
BIS electrode 1 is
detached from the patient.
1. Check whether BIS electrode 1 is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
2 High Imped.
Low
The impedance of BIS
electrode 2 is too high.
1. Check whether BIS electrode 2 is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
2 Lead Off
Low
BIS electrode 2 is
detached from the patient.
1. Check whether BIS electrode 2 is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
3 High Imped.
Low
The impedance of BIS
electrode 3 is too high.
1. Check whether BIS electrode 3 is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
3 Lead Off
Low
BIS electrode 3 is
detached from the patient.
1. Check whether BIS electrode 3 is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
4 High Imped.
Low
The impedance of BIS
electrode 4 is too high.
1. Check whether BIS electrode 4 is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
4 Lead Off
Low
BIS electrode 4 is
detached from the patient.
1. Check whether BIS electrode 4 is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
G High Imped.
Low
The impedance of BIS
electrode G is too high.
1. Check whether BIS electrode G is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
G Lead Off
Low
BIS electrode G is
detached from the patient.
1. Check whether BIS electrode G is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
Low
The impedance of BIS
1. Check whether BIS electrode C is
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...