8-70
8. Check the ACGO assembly.
Required tools:
One negative pressure ball
One test fixture of the circuit transit adapter
One PU tube (6X100)
Two 3106-08-10 adapter connectors
One 3126-06-00 tube plug
One 3126-08-00 tube plug
Two 3126-10-00 tube plugs
Test procedures:
(1) Disable the system switch.
(2) Remove PU tube 21 and PU tube 22 connecting to the ACGO assembly and circuit
adapter. Remove only the end connecting to the ACGO assembly.
(3) Block the end with the pipeline removed using two 3126-10-00 tube plugs.
(4) Perform steps 3 to 7 in 4 Leakage Test on All Pipelines in the Circuit Adapter. If the test
is not passed, the circuit adapter connector or sealing part is damaged. If leakage exists,
insert the removed pipeline to the ACGO assembly. Note that the position of the black
line on PU tubes 21 and 22 should be the same as that before the pipeline is removed.
(5) Remove the hose 52 and hose 53 used to connect the O2 flush assembly and the
vaporizer support assembly to the ACGO assembly. Remove only the end of the hoses
connecting to the ACGO assembly.
(6) Block the end of the ACGO assembly with the hoses removed using the 3126-06-00
tube plug and 3126-08-00 tube plug.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...