8-60
3.
Push the pushrod of the syringe to make the pressure reading on the anaesthesia machine
calibration apparatus rise to 70-90 cmH2O. Then, stop pushing. Remain the relative
position between the pushrod and the syringe body unchanged. If the pressure reading of
the anaesthesia machine calibration apparatus decreases no more than 5 cmH2O within
15s, the test item is passed.
If test item I is not passed but test item II is passed, the pressure sampling line of the flow sensor
in the breathing system is damaged. In this case, replace the breathing system. If both test item I
and test item II are not passed, check the internal sampling line and connector of the system and
the sealing part and solenoid valve of the circuit transit block assembly, until test item II is passed.
Then, conduct test item I. If test item I is still not passed, the pressure sampling line of the flow
sensor in the breathing system is damaged. In this case, replace the breathing circuit.
8.4.4.2 Leakage Test on the Low-pressure Airway System
After you confirm that the pressure sampling line of the flow sensor is not leaked, perform the
leakage test on the low-pressure airway system, as shown in Figure a to Figure d.
Figure a System leakage test
1. Leakage test on
the breathing
system in
mechanical
ventilation mode
.
Not pass
Pass
Pass
Pass
Not
pass
1. Leakage test
on the breathing
system in
mechanical
ventilation
mode
Pass
Figure d Leakage test on
the breathing circuit
Figure b Manual
ventilation
leakage test
4. Leakage
test on all
pipelines in
the circuit
adapter
Figure c
Leakage
test on the
internal
Re-install the bellow and
soda lime canister (clean
the
entrance
and sealing
washer
of
the canister),
and check the bellow
Completed
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...