4-2
4.2 System Self- Test
When the system is powered on, it performs a self-test to ensure its alarm system (alarm LED,
speaker, and buzzer) and hardware (flowmeter board, ventilator board, assistant ventilator board,
power board, and CPU board) are properly functioning.
System Self-test Sequence
Comments
1. A high-pitched beep is sounded. A click is sound if O
2-
N2O drive
switch is configured.
Alarm self-test
2. The startup screen is displayed.
/
3. The LED above the touchscreen illuminates in sequence: red, yellow,
and blue.
Alarm self-test
4. A test low priority alarm is sounded.
Alarm self-test; no
abnormal sound
5. The System Self-Test progress bar is displayed.
/
6. The System Self-Test is automatically started.
Hardware self-test
7. The results of the System Self-Test are displayed.
/
Ensure the system self-test is passed, and no abnormal alarm messages.
NOTE
Make sure that the breathing circuit is correctly connected and not damaged.No
abnormal sounds of buzzer, such as breaking sound, hoarse sound, changed sound,
sound too large, or sound too small, during the system self-test.
All needle valves are closed totally during the system self-test.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...