2-46
2.2.2 Pneumatic Components and Symbles
2.2.2.1 Pneumatic Components
The series numbers of components correspond with the number in the above pneumatic circuit
diagram.
SN Name
SN Name
1
O2 P- Line
37 Inspiratory flow sensor
2
O2 cylinder
38 Expiratory flow sensor
3
Air P- Line
39 O2 sensor
4 Air
cylinder
40 Scavenging reservoir and muffler
5
N2O P- Line
41 Manual/Auto switch
6
N2O cylinder
42 Manual bag
7
Regulator (0.4MPa)
43 APL valve
8
Pressure relief valve (0.758MPa)
44 Gas module
9 Filter
45 Bellows
assembly
10 Regulator
(0.2MPa)
46 Auxiliary
O2
supply
11 Inlet gas flow regulator
47 Airway pressure gauge
12 Flow sensor (Venturi)
48 Pressure sensor
13
Mechanical overpressure valve
(100cmH2O)
49
Water collection cup
14 Pop-Off
valve
50 Single-vaporizer
manifold
15 PEEP safety valve
51 Pressure relief valve (10cmH2O)
16 Pressure switch (125KPa)
52 Negative pressure valve (1cmH2O)
17 Proportional PEEP valve
53 Pressure sensor
18 Expiratory
valve
54 AGSS
19 Pneumatic resistor
55 Pressure relief valve (11KPa)
20 O2 flush valve
56 Venturi negative pressure generator
21 Flow
restrictor
57 Muffler
22 System switch
58 Adjustable negative pressure gauge
23 Pressure switch (0.2MPa)
59 Floating overfill protection valve
24 Regulator (0.2MPa)
60 Negative pressure suction filter
25 /
61 Liquid
bottle
26 /
62 Backup oxygen supply connector
27 /
63 Free
expiratory
valve
28 /
64 Drive
Gas
Switch
29 Double-vaporizer manifold
65 Backup oxygen flowmeter
30
Check valve
66
Two-position two-way switch valve
(NO)
31
/
67
Two-position two-way switch valv
(
NC
)
32
ACGO selector switch
68
Two-position three-way switch valve
(
NC
)
33 Inspiratory
valve
69 Proportional
valve
34 CO2
absorber
70 Oxygen
flow
sensor
35 BYPASS cut-off valve
71 Nitrous oxide and air flow sensor
36 Expiratory valve
72 Total flow sensor
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...