5-18
The preceding logs can be exported by choosing [
Export data
] in the [
System
] menu.
If the number of logs exceeds 500, the new one will cover the older one. And the data will
not be lost after powering off and on or upgrading software.
5.5 System Info
System information includes Software Version information, Running time, and Register.
Software version includes the joint version and the versions and release dates of the system
program, bootstrap, FPGA display program, FPGA sound program, VPM software, VCM software,
power board software, language, EFCS software, EFCS-PM software, and key board software, etc.
If the combination of all current software versions matches a joint version in the version file, the
joint version is displayed as the correct version.
If the combination of all current software versions does not match any joint version in the version
file, the joint version is displayed as Fail.
If the combination of all current software versions does not match any joint version in the version
file, the software version of the system program and the software versions of other modules not
matching the system program are displayed in red.
The running time includes the machine's running time and the ventilator's running time.
Register records data related to running exceptions of the machine.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...