4-27
NOTE
In case of leak test failure, check the machine for leakage and roughly assess the
amount of leakage by using the following methods.
Method 1: In the default VCV mode, stop fresh gas supply. If the bellows rises to
the top each time, it indicates that the machine is not leaky. Otherwise, the machine
is leaky. Gradually increase fresh gas. The amount of fresh gas when the bag rises
to the top at each expiration can be roughly calculated as the amount of leakage.
Method 2: During leak test, observe the airway pressure gauge. A period of time
(about 30 s) belongs to pressure holding stage after the airway pressure rises. If the
airway pressure gauge shows that airway pressure is gradually falling, it indicates
that the machine is leaky. Slowly increase fresh gas until airway pressure stops
falling. The amount of the then fresh gas can be calculated as amount of leakage.
If there is indeed a leak, check the pneumatic circuit system for leakage and
troubleshoot the problems as described in section 8.3 Leak and Compliance Tests.
After leak failure is troubleshot, do the leak test again and make sure the test is
passed.
4.11.5 APL Valve Test
1. Make sure that the system is Standby. If not, press the
key and select [
Ok
] from the
pop-up menu to enter Standby.
2. Set the Bag/vent switch to the bag position.
3. Connect the manual bag to the manual bag port.
4. Connect the Y piece on the breathing tube to the leak test plug on the manual bag port.
5. Turn the APL valve control to let the pressure of APL valve stay at 30 cmH
2
O.
6. Push the O2 flush button to inflate the manual bag.
7. Make sure that the reading on the airway pressure gauge is with the range of 25 to 35
cmH
2
O.
8. Turn the APL valve control to the SP position.
9. Set the O
2
flow to 3 L/min. Turn any other gases off.
10. Make sure that the reading on the airway pressure gauge is less than 5 cmH
2
O.
11. Push the O
2
flush button. Make sure that the reading on the airway pressure gauge does
not exceed 10 cmH2O.
12. Turn the O
2
flow control to set the O
2
flow to minimum. Make sure that the reading on
the airway pressure gauge does not decrease below 0 cmH
2
O.
4.11.6 O2 Sensor Related Tests
4.11.6.1 O2 Sensor Disconnection Alarm Test
1. Select
, and then [
System
], and enter password: 799789.
2. Set
[
O
2
Cell Monitoring
] to [
On
], and select [
Confirm
].
3. Disconnect the plug of O2 sensor, a prompt message of O2 Sensor Disconnected should
appear. Connect O2 fixture (0611-K08-XX), this message should disappear.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...