8-7
Message
Alarm
Priority
Cause
Solution
normally.
4. Replace the BIS module.
CO2 Self Test
Error
CO2 Self Test:
Time out
Low
An error occurs during the
CO2 module self test.
1. Reconnect the CO2 module, and
restart the machine for a self test again.
2. Check the cable connection between
the module rack and CPU board.
3. Check whether the module rack runs
normally.
4. Replace the CO2 module.
8.2.2 CPU Board Runtime Alarms
Message
Alarm
Priority
Cause
Solution
IP Address
Conflict
Medium
The IP address of
the machine is the
same as the IP
address of another
device in the local
network.
1. Re-configure the IP address.
2. If the problem persists, upgrade the system
software or replace the CPU board.
Fan Failure
Medium
The speed of the fan
is lower than the
rated speed by 20%.
1. Check whether the fan stops running, or the
speed is low (the normal speed is about 4000
RPM).
2. Reconnect the fan cable.
3. If the problem persists, check whether the
supply voltage of the 12 V fan is within the
valid range. If not, check the power board.
4. If the problem persists, replace the fan.
5. If the problem persists, replace the CPU
board.
Fan Failure 02
Medium
The speed of the fan
in the module rack
is lower than 3640
RPM.
1. Check whether the fan stops running, or the
speed is low (the normal speed is about 4000
RPM).
2. Reconnect the fan cable.
3. If the problem persists, check whether the
supply voltage of the 12 V fan is within the
valid range. If not, check the power board.
4. If the problem persists, replace the fan.
5. If the problem persists, replace the CPU
board.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...