8-79
5. For any sampling line connection errors, reconnect the sampling line and ensure that it is
connected correctly.
Perform the following operations to diagnose the measurement error of the pressure
sensor:
As the actual pressure increases, the measurement value of the pressure sensor
increases accordingly. Otherwise, the calibration data is incorrect. In this case, you
need to recalibrate the pressure sensor.
The measurement value of the pressure sensor is accurate compared with that of
standard pressure measurement equipment (calibration equipment of the
anaesthesia machine). Otherwise, the calibration data is incorrect. In this case, you
need to recalibrate the pressure sensor.
For details, see section 4.14.3 Constant Pressure Test (Checking the Pressure Sensor
Accuracy).
8.5.7 Diagnosing Inspiratory valve Exceptions
You can check whether the status of the inspiratory valve is correct by using the valve diagnosis
tool.
1. The tubes of the anaesthesia machine are connected according to the settings for
checking whether the sampling line of the pressure sensor is connected normally. For
details, see section 8.5.3 Preparations Before Using the Valve Diagnosis Tool.
2. In
the
Diagnostic Tests
menu, increase the D/A value of the inspiratory valve gradually.
If the measurement values of the embedded flow sensor, inspiratory flow sensor and
expiratory flow sensor change slightly and the air flow at the interface of the water
collection cup in the breathing system is light, the inspiratory valve or VCM has D/A
problems.
3. In normal cases, if the D/A value is set to 2500 for the inspiratory valve, the flow
volume measured by the standard flow measurement equipment can reach 90 L/min.
4. If the D/A value is set to be over 4000 for the inspiratory valve, the flow volume
measured by the standard flow measurement equipment cannot reach 90 L/min and the
flow calibration fails. In this case, you need to replace the expiratory valve or VCM.
5. To check whether the VCM has DA output problems, measure the DA output
corresponding to the inspiratory valve on the VCM by using a multimeter. If the voltage
increases as the DA increases and approaches 6 V when the DA is over 4000, the DA
corresponding to the inspiratory valve of the VCM may be normal.
6. After replacing the expiratory valve or VCM, check whether the problem is solved by
using similar methods.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...