7-54
5. A screen as shown in the following figure is displayed if the ongoing calibration is
cancelled. Select [
Try Again
] to re-conduct calibration. Select [
Done
] to quit
calibration.
7.3.7.4 Common Failures and Recommended Actions
Failure Description
Possible Cause
Recommended Action
After [
Begin
] is
selected, a calibration
failure message is
displayed very soon.
The machine is in the
BFCS state.
Set the machine to the EFCS state.
The pressure of the
N2O supply is low.
Replace or connect to the gas supply to ensure
that the N2O pressure is within the specified
range.
The calibration fails
about 1 minute after
[
Begin
] is selected.
The calibration points
do not meet the
monotonic property.
Recalibration is recommended.
7.3.8 CO2 Calibration (factory)
7.3.8.1 Preparations
Prepare the following before doing the calibration:
Gas cylinder: one or more cylinders filled with 3% , 4%, 5%, 6% , or 7% CO
2
T-shape connector
Samping line
7.3.8.2 Calibration Procedures
NOTE
During the calibration, selecting [
Calibrate
] again does not take effect or exit the
calibration menu. Other operations than menu options are disabled until the end of
calibration.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...