8-3
Message
Alarm
Priority
Cause
Solution
3. An error occurs during the
self test of the CPU, RAM,
address line, watchdog, Flash,
O2/N2O/AIR proportional
valve, or FPGA.
3. If the problem persists, replace the
CPU board.
Flowmeter
Voltage Error
High
The sample values of the
DVDD, AVDD, or VC voltage
of the current flowmeter are
not in the normal range.
Restart the machine for a self test
again.
Aux Control
Module Self Test
Error
Aux Control
Module Self
Test: Time out
High
1. CPU, Flash or WTD error.
2. After the machine is
restarted, the CPU board
cannot communicate with the
Aux Control board.
1. Restart the anaesthesia machine.
2. Reconnect or replace the
communication line between the CPU
board and the Aux Control board.
3. If the problem persists, replace the
ventilation protection board.
4. If the problem persists, replace the
CPU board.
Ventilator Self
Test Error
Ventilator Self
Test: Time out
High
1. CPU, TIMER, RAM, WTD,
EEPROM or AD error
2. After the machine is
restarted, the CPU board
cannot communicate with the
VCM board.
1. Restart the anaesthesia machine.
2. Reconnect or replace the
communication line between the CPU
board and the VCM board.
3. If the problem persists, replace the
VCM board.
4. If the problem persists, replace the
CPU board.
Ventilator
Voltage Error
High
5 V or 12 V voltage error
1. Restart the anaesthesia machine.
2. Measure the supply voltage of the
VCM board and check whether the
voltage meets the requirement.
3. If not, replace the power board or
cable.
4. If the problem persists, replace the
VCM board.
PEEP Valve
Failure
Medium
1. PEEP valve voltage error
2. PEEP valve pressure error
1. Check whether the accuracy of the
PEEP sensor is within the valid range.
On the [Maintenance] menu, calibrate
the sensor or replace VCM.
2. Test the voltage at the corresponding
testing point.
3. Check the installation of and
connection between the power cable
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...