1-2
1.3 Cautions
CAUTION
This device uses high pressure compressed gas. When attaching or disconnecting
backup gas cylinders, always turn the cylinder valves slowly. Use the flow meters to
bleed down the pressure, watching the cylinder gauge indicate the depleting
cylinder pressure, before disconnecting the cylinder from the yoke. Always open
and close cylinder valves fully.
This device operates using compressed gas at high pressures from the hospital
central supply. When connecting gas supply lines attach the hose connection to the
machine before connecting the quick disconnect fitting to the hospital source.
Disconnect the supply hose from the hospital source connection prior to
disconnecting it from the gas connection fittings.
Refer to the 6.2Maintenance Period for assistance when performing scheduled
periodic maintenance.
Do not leave gas cylinder valves open if the pipeline supply is in use and the system
master switch is turned to 'ON'. If used simultaneously, cylinder supplies could be
depleted, leaving an insufficient reserve supply in the event of pipeline failure.
Use cleaning agent sparingly. Excess fluid could enter the machine, causing
damage.
This machine must only be operated by trained, skilled medical staff.
Perform the electrical safety inspection as the last step after completing a repair or
after routine maintenance. Perform this inspection with all covers, panels, and
screws installed.
After changing the CO2 absorbent, carry out a system leak test.
Only Selectatec™ compatible vaporizers with Interlock-System may be used with
the A5 unit.
After each exchange of a vaporizer, carry out a system Leak test.
Do not clean the machine while it is on and/or plugged in.
Pressing “cancel” at any time during the procedure will cancel the session's
settings and reload the previously-stored calibration coefficients.
Depleted soda lime changes color. Replace the soda lime if approximately 2/3 of the
absorber content is discolored. CO2 absorbent can be safely changed without
stopping mechanical ventilation.
This equipment contains parts which are easily damaged due to electrostatic
discharge (ESD). Follow ESD prevention program when touching, taking out, or
inserting parts or components.
1.4 Notes
NOTE
Unauthorized servicing may void the remainder of the warranty. Check with the
factory or with a local authorized distributor to determine the warranty status of a
particular instrument.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...