2-31
2.1.6.2 Monitoring Module
Monitoring module monitors pressure and flow of anesthesia machine and breathing system,
controls valves, monitors and collects statuses, reads O2 concentration, monitors pressure and
flow in circuit, and control accuracy accurately. Monitoring module contains VCM and VPM.
VCM monitrs the parameters including inspiratory flow, expiratory flow, interior flow of the
machine, airway pressure, PEEP, O2 concentration, etc, and controls the actions of three-way
valve and exhalation valve (including safety valve, PEEP valve, and inspiration valve).
VPM monitors the signal of the pressure switch of drive gas, ACGO limit switch, O2 pressure
switch at gas source inlet, auto/manual switch, CO2 absorbent canister in-place switch, etc. VPM
also controls the selector valve of drive gas, and controls safety valve with VCM, and send
monitoring information to the main board through series port.
J1, power supply
port and VCM
communication port
J5,VPM
communication port
J7,
communication
port of
calibration
J10,CO2
Absorber
canister
switch signal
J4,
exhalation
valve drive
J8, on/off
signal
J15, control
selector valve
of drive gas
J2,
Exhalation
valve Power
supply
J12, O2
sensor port
J6, three-
way valve
drive
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...