3-1
3
Installation Guide
3.1 Preparation
Before installation, contact the consumer and tell him/her to provide the following materials. The
customer is responsible for supplying these materials. Missing items may result in delays,
incomplete installations, and/or additional service visits.
Gas supply hose connector.
O2 cylinder, N2O cylinder, and Air cylinder in compliance with the configuration
standard.
Anesthetic vaporizer and adapter (if not purchased together with the anesthesia system)
Anesthetic agent
Loose fill or Pre-pak CO2 absorbent
Effective O2, N2O, and Air gas sources (280 to 600kPa (40 to 87psi))
Dropdown hoses for ceiling mounted medical gas equipment, compatible with
quick-disconnect hoses ( if not purchased together with the anesthesia system)
3.2 Assembly
NOTE
After the flow sensors are calibrated, the anesthesia machine must be used together
with the breathing system module. If the breathing system module is removed,
make sure that it is re-installed onto the original anesthesia machine. If other
breathing system module is to be installed, the flow sensor must be re-calibrated
(please refer to 7.3.2Flow calibration (Service))
.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......