7-45
7.3.6 O2 Sensor Calibration
NOTE
The O2 sensor needs to be calibrated when a great deviation occurs in the O2
monitored value or the O2 sensor or VCM is replaced.
Observe whether the O2 sensor displays values on the measure screen before
calibration. If not, check whether the O2 measure switch is turned on, check the
connection cables of the O2 sensor, or replace the O2 sensor until values are
displayed.
The O2 sensor needs to be calibrated at 21% O2 and 100% O2 and corresponding
standard O2 concentration gases are required.
7.3.6.1 21% O2 Calibration
Perform 21% O2 sensor calibration as follows:
1. Select
[
Main
] > [
General
] > [
Calibrate O2 Sensor
], or select [
Main
] > [
System
] >
[
Calibrate
] > [
O2 Sensor
] or [
Main
] > [
Service
] > [
Calibration
] > [
O2 Sensor
] to go
to the screen as shown in the following figure. The [
General
] tab page displays only the
21% O2 sensor calibration option. The [
System
] and [
Maintenance
] tab pages require
the password for 21% and 100% O2 sensor calibration. You can set the machine based
on the instructions displayed on the screen. Select [
Start
] to start calibration.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...