4-30
4.12.2 Test the O
2
Concentration Monitoring and Alarms
NOTE
This test is not required if O
2
sensor is not configured.
1. Set the Bag/vent switch to the
position.
2. Remove the O
2
sensor and make sure that the sensor measures approximately 21% O
2
in
room air.
3. Select
[
Alarm
] and then [
Limit
]. Set the FiO
2
low limit to 50%.
4. Make sure that a low FiO
2
alarm occurs.
5. Set the FiO
2
low limit to a value less than the measured FiO
2
value and make sure that
the alarm cancels.
6. Put the O
2
sensor back in the circuit.
7. Select
[
Alarm
] and then [
Limit
]. Set the FiO
2
high limit to 50%.
8. Connect the manual bag to the manual bag port. Push the O
2
flush button
to fill the
manual bag. After two to three minutes, make sure that the sensor measures
approximately 100% O
2.
9. Make sure that a high FiO
2
alarm occurs.
10. Set the FiO
2
high limit to 100% and make sure that the alarm cancels.
4.12.3 Test the Low Minute Volume (MV) Alarm
1. Make sure that MV alarm is turned on.
2. Select
[
Alarm
] and then [
Limit
]. Set the MV low limit to 8.0 L/min.
3. Make sure that a low MV alarm occurs.
4. Select
[
Alarm
] and then [
Limit
]. Set the MV low alarm limit to the default.
4.12.4 Test the Apnea Alarm
1. Connect the manual bag to the manual bag port.
2. Set the Bag/vent switch to the
position.
3. Turn the APL valve control to set the APL valve to the minimum position.
4. Inflate the manual bag to make sure that a complete breathing cycle occurs.
5. Stop inflating the manual bag and wait for at least 20 seconds to make sure that the
apnea alarm occurs.
6. Inflate the manual bag to make sure that the alarm cancels.
4.12.5 Test the Sustained Airway Pressure Alarm
1. Connect the manual bag to the manual bag port.
2. Turn the O
2
flow control to set the O
2
flow to minimum.
3. Turn the APL valve control to set the APL valve to 30 cmH
2
O position.
4. Set the Bag/vent switch to the
position.
5. Push the O
2
flush button for approximately 15 seconds. Make sure that the sustained
airway pressure alarm occurs.
6. Open the patient connection and make sure that the alarm cancels.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...