7-5
7.3.1.1 Flow Calibration Principles
This calibration item is used to calibrate only the flow sensors and inspiration valve in the circuit.
The built-in flow measurement base source of the machine is utilized to calibrate the inspiratory
flow sensor and expiratory flow sensor on the breathing system. The VCM opens the inspiration
valve based on inspiration valve DA values obtained from service calibration, to inflate the circuit
of the breathing system, records the flow of the built-in flow sensor, and uses the measured value
of the built-in flow sensor as well as AD values collected by the inspiratory flow sensor and
expiratory flow sensor as the data of one calibration point for flow sensors, and uses the measured
value of the built-in flow sensor as well as the DA value of the inspiration valve as the data of one
calibration point for the inspiration valve. Change the DA value of the inspiration valve to obtain
calibration data under a series of flows, forming a flow calibration data table. The flow should be
in the range from 0 L/min to 120 L/min when the inspiration valve is opened during flow
calibration.
If the inspiratory flow sensor and expiratory flow sensor have been used for months, for example,
three months after calibration, the measured tidal volume greatly deviates from the settings (with
the deviation of more than 9%) due to sensor aging or environmental factors; or a flow sensor is
replaced. In this case, the flow sensor needs to be calibrated. The flow calibration (user) can be
adopted.
FIGURE 1 Schematic Diagram of Flow Calibration (User)
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...