2-12
Pin
No.
Signal
Name
Discription
16)
NTC2
Lithium battery 2, connected with thermistor signal
inside
17)
BAT1+
Lithium battery 1 input, connected to the anode of
battery
18)
GND
Ground
19)
BC1
Lithium battery 1 in-place signal. Low level indicates
there is battery; high level indicates there is no battery
20)
NTC1
Lithium battery 1 connected with thermistor signal
inside
21)
GND
Ground
22)
GND
Ground
23)
P15V
Heating wire drive voltage output
24)
P15V
Heating wire drive voltage output
25)
Temperature-R12
Thermistor pin 1 of heating wire
26)
Temperature-R22
Thermistor pin 1 of heating wire
27)
Temperature-R11
Thermistor pin 2 of heating wire
28)
Temperature-R21
Thermistor pin 2 of heating wire
29)
GND
Ground
30)
GND
Ground
31)
3V3
3.3V power supply voltage output
32)
3V3
3.3V power supply voltage output
33)
5V
5.0 V power supply voltage output
34)
GND
Ground
35)
GND
Ground
36)
GND
Ground
37)
12VB
Second 12V power supply voltage output
38)
GND
Ground
39)
12VB
Second 12V power supply voltage output
40)
12VB
Second 12V power supply voltage output
41)
GND
Ground
42)
GND
Ground
43)
12VA
First 12V power supply voltage output
44)
GND
Ground
45)
12VA
First 12V power supply voltage output
46)
12VA
First 12V power supply voltage output
47)
NC
No connection inside
48)
15.2V
15.2V power supply voltage output
49)
NC
No connection inside
50)
15.2V
15.2V power supply voltage output
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...