2-13
Definition of mother board J11 (main control board port)
Pin No. Signal Name
Discription
1)
LCD3V3
LCD power supply
2)
GND
Ground
3)
NC
No connection inside
4)
NC
No connection inside
5)
RXD_HW_OR_JH
Received signal of infrared backpanel or patient monitor
6)
TXD_HW_OR_JH
Sent signal of infrared backpanel or patient monitor
7)
GND
Ground
8)
RXD_AG_232
Received signal of interior AG module
9)
TXD_AG_232
Sent signal of interior AG module
10)
GND
Ground
11)
Touchpad_5V
5V power supply of touchpad
12)
Touch
Touchpad USB data
13)
Touchpad_usb-
Touchpad USB data signal-
14)
GND
Ground
15)
SCL_CPU
Alarm light board IIC clock signal
16)
SDA_CPU
Alarm light board IIC data signal
17)
GND
Ground
18)
MAIN_ACGO_CTRL2
ACGO control signal 2
19)
NC
No connection inside
20)
ACGO_State1
ACGO
status signal 1
21)
NC
No connection inside
22)
FAN_State2
Fan status signal 2
23)
FAN_State1
Fan status signal 1
24)
GND
Ground
25)
3V3
3.3V power supply of main control board
26)
3V3
3.3V power supply of main control board
27)
GND
Ground
28)
GND
Ground
29)
5V
5V power supply of main control board
30)
5V
5V power supply of main control board
31)
GND
Ground
32)
GND
Ground
33)
NC
No connection inside
34)
NC
No connection inside
35)
NC
No connection inside
36)
SPEAK-
Speaker drive
37)
SPEAK+
Speaker drive signal -
38)
GND
Ground
39)
VGA_Red
VGA red signal
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...