4-41
4.14.3 Constant Pressure Test (Checking the Pressure Sensor
Accuracy)
NOTE
Generally, measurement deviations seldom occur to pressure sensors. However, in
case that the ventilator control board, solenoid valve assembly, or expiratory valve
assembly needs to be replaced, you must perform pressure calibration and check
the pressure sensor accuracy so as to confirm the effectiveness of calibration.
You can use any flowmeter with an accuracy of at least
±
2% for the accuracy
measurement of the pressure sensors.
To check the measurement accuracy of pressure sensors, perform the following operations:
1. Complete pneumatic connections as follows:
The following picture shows a four-way device.
a. Remove the top cover (3 captive screws).
b. Remove the two tubes marked as #72 and #9 from the pressure sensors (refer to a figure
below).
c. Connect the four-way tube to the pressure sensor P1 of the monitor board, PEEP
pressure sensor P2, the tube marked as #72, and the low pressure port of Fluke VT-Plus.
The tube marked as #9 will remain unconnected for this calibration.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......