7-36
Failure
Description
Possible Cause
Recommended Action
switch can be set normally.
[00000008] is
displayed.
There is a zero point error for the
flow pressure sensor.
1. Check the zero point.
2. Replace the VCM.
[00000010] is
displayed.
There is a zero point error for the
PEEP pressure sensor.
1. Check the zero point.
2. Replace the VCM.
[00000020] is
displayed.
The measured range of the flow
pressure sensor is abnormal.
1. Check the sampling line connection
and tightness.
2. Check the gas supply pressure.
3. Check the configuration of the
calibration device.
4. Troubleshoot the sensor and valve
fault using the valve diagnosis tool.
For details, see section 5.4.
5. Replace the VCM.
[00000040] is
displayed.
The measured range of the PEEP
pressure sensor is abnormal.
1. Check the sampling line connection
and tightness.
2. Check the gas supply pressure.
3. Check the configuration of the
calibration device.
4. Troubleshoot the sensor and valve
fault using the valve diagnosis tool.
For details, see section 5.4.
5. Replace the VCM.
[00000080] is
displayed.
The calibration data of the flow
pressure sensor does not satisfy
monotonicity.
1. Check the sampling line connection
and tightness.
2. Check the gas supply pressure.
3. Troubleshoot the sensor and valve
fault using the valve diagnosis tool.
For details, see section 5.4.
4. Replace the VCM.
[00000100] is
displayed.
The calibration data of the PEEP
pressure sensor does not satisfy
monotonicity.
1. Check the sampling line connection
and tightness.
2. Check the gas supply pressure.
3. Troubleshoot the sensor and valve
fault using the valve diagnosis tool.
For details, see section 5.4.
4. Replace the VCM.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...