4-34
10. Verify that the tidal volume displayed on the anesthesia machine is within 7% (±42 mL)
of the set value within approximately the first 1 minute after ventilation starts.
11. Verify that the measured O2 concentration is at least 97% after 5 minutes
12. Set the air flow to 3 L/min and disable the O2 flow rates.
13. Verify that the measured O2 concentration is 21% ±3% (volume fraction) after 5
minutes.
4.13.4 VCV Child Ventilation Mode Test
1. Attach a breathing circuit and breathing bag.
NOTE
For testing purposes always use a reusable breathing circuit.
2. Attach an adult test lung to the Y-piece of the breathing circuit.
NOTE
Limit the volume in the test lung to provide sufficient airway pressure to satisfy the
Low Peak Pressure alarm. Or set the peak pressure alarm limit to a smaller value
to prevent the alarm when an adult test lung is used.
3. Attach a ventilation tester between the expiratory port and the breathing hose.
4. Select Air in [Balance Gas] and single-tube flow in [Fresh Gas Control]. Set the O2 flow
to 2 L/min and set the N2O and air flow rates to the minimum value.
5. Set breathing parameters under standby mode.
Parameter
Setting
Patient Size
Child
Ventilation Mode
VCV
Vt 120
Rate 15
I:E 1:2
Tpause 10
PEEP Off
Plimit 40
6. Press the [Set Mode] button to exit standby mode and begin ventilation.
7. Verify that the pressure waveform, respiratory capacity, mean or plateau pressure,
respiratory rate, and minute ventilation appear on the screen.
8. Verify that the displayed tidal volume is within 15 mL of the set value within
approximately the first 1 minute after ventilation starts.
9. Verify that the tidal volume, measured by a ventilation tester at the expiratory port, is
within 15 mL of the set value within approximately the first 1 minute after ventilation
starts.
4.13.5 Air Way Disconnection Alarm Test
1. Disconnect the expiratory limb from the expiratory port of breathing system when the
ventilator is working.
2. Ensure
the
following alarms appear:
A low Paw alarm appears on the screen
Alarm sounds
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...