392
Error Messages
Msg
No.
Message
Remedy
References
57
Cannot set.(External sampling)
Cannot use roll mode when using
external sampling.
58
Rated capacity/rated output error. Rated capacity/Rated output has
exceeded selected range. Input
correct value.
59
Cannot use. (Roll Mode, Wave
Calculation)
If the memory division function is
in use, these functions cannot be
used.
60
No waveform data.
Acquire waveform data.
—
68
Recording length is too long.
Shorten the measurement
recording length.
The maximum recording length
that can be calculated is 40,000
div.
“3.3.3 Recording Length (Number
80
In key lock.
Keylock is engaged. Cancel the
Key-Lock function.
91
LAN: Bad IP address.
Check the IP address.
“16.1 LAN Connection and Settings
95
LAN: Connection timed out.
Check the transmission settings.
97
LAN: Network error.
Check the connection between
the instrument and connection
destination.
100 Inappropriate setting of time axis
for waveform data.
Return the timebase to the
timebase setting when the
waveform was measured.
101 Some blocks are different in time
axis setting from the others.
Set the same timebase setting
for all search target blocks.
If they cannot be the same,
set the search range to display
blocks only.
102 Some blocks are different in
modules configuration from the
others.
Unify the modules configuration
of search target blocks.
If they cannot be the same,
set the search range to display
blocks only.
103 No data matching your search
condition.
Check the trigger setting.
104 There is no data in the search
target channel.
Select channels with
measurement data in the search
target.
105 There is a block with no data in
the search target channel.
Select blocks with measurement
data in the search target.
106 Some blocks are different in
measurement mode of modules
from the others.
Blocks with varying measurement
modes cannot be searched.
—
108 Some blocks are different in
measurement mode of modules
from the others.
Set the search range within the
display blocks.
110 No space to store waveform data
at copy location.
Create empty space by deleting
waveform data from copy
location.
111 No space to register arbitrary
waveform data.
Create empty space by deleting
waveform data from registration
location.
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......