373
Module Specifications
18.6.10 Model U8974 High Voltage Unit
The accuracy is provided for the condition where the module is installed on the Memory HiCorder at 23°C±5°C
(73°F±9°F), 20% RH to 80% RH, and zero adjustment is executed 30 minutes after power on.
Product warranty
period
3 years
Guaranteed accuracy
period
1 year
Number of input
channels
2 channels
Measurement
functions
Instantaneous value, r.m.s. value (can be changed for each channel)
Measurement range
200 mV, 500 mV, 1 V, 2 V, 5 V, 10 V, 20 V, 50 V/div (mode DC)
500 mV, 1 V, 2 V, 5 V, 10 V, 20 V, 50 V/div (mode RMS)
Measurement
accuracy
±0.25% f.s. (Filter 5 Hz On)
RMS measurement
accuracy
±1.5% f.s. (DC, 30 Hz to 1 kHz, sine waves, response time: slow)
±3% f.s. (1 kHz to 10 kHz, sine waves)
Crest factor: 2 (sine waves, the peak voltage is up to 1000 V)
RMS measurement
response time
Fast: 150 ms (rising from 0% f.s. to 90% f.s.)
Normal: 500 ms (rising from 0% f.s. to 90% f.s.)
Slow: 2.5 s (rising from 0% f.s. to 90% f.s.)
Temperature
characteristic
±0.05% f.s./°C
Frequency
characteristic
DC to 100 kHz −3 dB
Noise
30 mV p-p (typ.), 50 mV p-p (max.) (highest sensitivity range, with input shorted)
Common mode
rejection ratio
80 dB or more (50 Hz/60 Hz input shorted)
Low-pass filter
OFF, 5±50%, 50±50%, 500±50%, 5 k±50%, 50 k±50% (Hz) -3 dB
Input type
Unbalanced input (floating)
Input coupling
DC/GND
Input resistance
4 M
Ω
±1%
Input capacitance
5 pF or less (at 100 kHz)
A/D resolution
16 bits
Maximum sampling
rate
1 MS/s
Input terminals
Banana jacks
Maximum input
voltage
1000 V DC, 700 V AC
Maximum rated
voltage to earth
1000 V AC/DC measurement category III, 600 V AC/DC measurement category IV
(between each input channel and chassis, between input channels)
Anticipated transient overvoltage: 8000 V
Operating
temperature and
humidity
As per Memory HiCorder in which Model U8974 is installed
Operating
environment
As per Memory HiCorder in which Model U8974 is installed
Specifications
18
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......