Connect to External Control Terminals
17.1 Connect to External Control Terminals
The method for connecting to the external control terminals is as follows.
Procedure
Cables to connect
Solid wire
Stranded wire
10 mm
Recommended wire
Solid wire: 0.65 mm dia. (AWG22)
Stranded wire: 0.32 mm
2
(AWG22)
Compatible wire:
Solid wire: 0.32 to 0.65 mm dia. (AWG28 to 22)
Stranded wire: 0.08 to 0.32 mm
2
(AWG28 to 22)
Strand diameter: 0.12 mm or more (per wire)
Stripped length:
9 to 10 mm
Button pressing tool:
Flat-blade screwdriver (shaft diameter: 3 mm, tip
width: 2.6 mm)
Connection procedure
Back side
1
2
3
1
Push in the button on the connector with
a flat-blade screwdriver or other tool.
2
With the button held in, insert the cable
into the cable connection hole.
3
Release the button.
The cable is locked.
Terminal
No.
Operation
1
Input external signal and execute the
following
•Start/end measurement
•Print/save data
•Pen Up/Down Operation (for X-Y
recorder operation)
2
3
4
—
5
Output instrument status as a signal
6
7
—
8
Input external signal and set sampling
rate
9
—
10
Output signal when triggering occurs
11
Input external signal as trigger source
17
External Control
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......