201
Pre-trigger Settings
9.8 Pre-trigger Settings
This applies to the Memory and FFT functions only.
By setting a portion (number of divisions or percentage) of the recording length to occur before triggering,
the waveform can be recorded before as well as after the trigger point.
You can also set the duration of a waveform to be recorded after a trigger point.
T
Pre-Trigger Setting
Amount
Recording
Length
T
Pre-Trigger Setting
Amount
Recording Before Triggering
Recording for a Specified Duration
After Triggering
Recording
Length
•
When all trigger sources (analog, timer trigger, etc.) are disabled (Off), pre-trigger settings are
ignored.
•
Pre-triggers cannot be set by div with the FFT function.
9.8.1
Set the Trigger Start Point (Pre-trigger)
Set the position of the trigger point relative to the specified recording length.
There are two setting methods. (Setting by percent and setting by div)
Procedure
Press the
DISP
key to open the Waveform screen.
1
Select the unit to use when setting the pre-trigger.
Move the cursor to
[Pre-trigger].
%
Set the pre-trigger in %.
(Default setting)
div
Set the pre-trigger in number of divs. For external
sampling, set by number of samples.
2
Enter the numerical value.
(When
[%]
is selected)
With the recording start point as 0% and the recording end
point as 100%, set the % for the trigger point.
-95%, 100%, 95%, 90%, 80%, 70%, 60%, 50%, 40%, 30%, 20%,
10%, 5%, 2%, 0%
(When
[div]
is selected)
With the recording start point as 0 div and the recording end
point as the set recording length, set the number of div for the
trigger point.
-Recording length (div) to +Recording length (div)
9
Trigger Settings
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......