109
Auto-print Settings
6.2 Auto-print Settings
This applies to the Memory, Recorder, and FFT functions.
These settings must be set before starting measurement.
Measurement data will print automatically when you press the
START
key to start measurement. For
printer output, make sure that the recording paper is loaded correctly. For LAN output, make sure that the
PC connected via LAN is prepared.
Procedure
Press the
SYSTEM
key to open the System screen, and then select the
[Printer]
sheet.
1
Enable auto-print.
(When using the Memory or FFT Functions)
Move the cursor to
[Auto Print]
, and select
[On]
.
Default setting: Off (no automatic printing)
(When using the Recorder Function)
Move the cursor to
[Realtime Print]
.
Select
[On]
.
•
To output waveform data to
the PC connected via LAN
Model 9333 LAN
Communicator must be
installed on the PC to which
the data is to be output.
For more information
about how to operate and
configure the Model 9333
LAN Communicator, as well
as install it on a PC, see the
instruction manual of Model
9333 LAN Communicator.
•
When setting the output
destination to
[LAN]
The IP addresses of the
instrument and the PC
connected via LAN must be
set. For more information
about how to set the IP
2
(When using the Memory or FFT Functions)
Specify the output destination.
Move the cursor to
[Output]
.
Printer:
LAN:
The data is automatically output to the PC connected
via LAN.
Model 9333 LAN Communicator required.
Printer
The data is automatically printed by the internal
printer. (
Default setting)
LAN:
The data is automatically output to the PC
connected via LAN.
Model 9333 LAN Communicator required.
3
(When using the Memory Function)
Choose print area settings as required.
Move the cursor to the
[Print Area]
item.
Whole Wave
Print all waveform data in internal memory of the
unit. (
Default setting)
A-B Wave
Print only A and B cursors defined range of
waveform data in the internal memory of the unit.
With the Recorder function, the whole waveform will be printed
during measurement regardless of the print area setting.
6
Printing (
When Model U8350 Printer Unit is installed
)
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......