377
Module Specifications
Sweep Function Specifications
Sweep waveforms
Non-DC FG waveforms, arbitrary waveforms
Sweep form
Linear
Sweep target
FG waveforms:
Frequency, amplitude, offset, duty ratio (pulse waves only)
(Simultaneous sweeping of frequency, amplitude, and offset)
Arbitrary waveforms: Clock frequency, amplitude, offset
(Simultaneous sweeping of clock frequency, amplitude, and offset)
Sweep time setting
range
10 μs to 1000 s (setting resolution: 10 μs)
Program Function Specifications
Sequence length
Output by connecting up to 128 steps
Step control
Ability to set an FG waveform, sweep waveform, or user waveform for each step
Ability to set the number of loops (for sweep waveforms) or output time (for FG and
arbitrary waveforms) for each step
Hold settings
ON/OFF setting for each step
Output time setting
range
10 μs to 1000 s (FG waveforms, arbitrary waveforms)
Step loop count
setting range
1 to 1,000 (for sweep waveforms)
Overall loop count
setting range
1 to 50,000 or ∞
Monitor functions
Display of step number being executed, step loop count, and overall loop count
Other Specifications
Channel
synchronization
Ability to set the phase between channels on the same module and between modules
Self-test function
Ability to monitor output voltage value
Monitor resolution: 10 mV
Monitor accuracy: ±3.0% f.s. (f.s.=15 V)
Output start/stop
By operating key on the Memory HiCorder or applying a signal to the unit’s external control
terminal
External input
Cancel hold and transition to next step on low-level signal input from an external source
while using the program function.
Control voltage level:
3.5 V to 5.0 V (high level), 0 V to 0.8 V (low level)
Response pulse width:
100 μs or greater (low level)
External output
Output during waveform output
Output format:
Open-drain output (with 5 V voltage output, active low)
Output voltage level:
4.0 V to 5.0 V (high level), 0 V to 0.5 V (low level)
Maximum switching capacity: 5 V DC to 30 V DC, 50 mA
External I/O
terminals
Push-button type terminal block
Waveform output
indicator
Red LED on during waveform output and off when output is off
Specifications
18
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......