194
Analog Signal Triggers
About period range settings
The period range settings for period triggering depend on the sampling period (sampling rate). (Changing
the timebase also changes the period setting range.)
The [Sampling Rate] setting can be verified on the Status screen -
[Status]
sheet.
The upper threshold of the period range cannot be set below the lower threshold, and vice-versa.
Lower threshold: can be set either to zero, or to at least five times the sampling period.
Upper threshold: can be set to no more than 20,000 times the sampling period.
To apply a trigger when the frequency increases (shorter period) above the upper threshold
Set the period trigger type to
[Per.I]
and the lower threshold to
[0]
. The lower threshold will be ignored
and triggering will occur when the frequency exceeds the upper threshold.
To apply a trigger when the frequency decreases (longer period) below the upper threshold
Set the period trigger type to
[Per.O]
and the lower threshold to
[0]
. The lower threshold will be ignored
and triggering will occur when the frequency drops below the upper threshold.
About the trigger point of the Out-of-Period trigger
Triggering occurs when the period of sequential crossings of the specified reference voltage exceeds the
period range.
The point at which triggering occurs depends on the specified period range and the period of the
measured signal.
When the input signal period is shorter than the specified lower threshold (Trigger Slope: ↑)
Reference
Voltage Level
Lower Threshold
Input Signal
Period
T
Triggering occurs when the rising edge (↑ trigger slope) of
the input signal crosses the reference voltage level, before
reaching the lower period threshold.
When the input signal period is longer than the specified upper threshold (Trigger Slope: ↑)
Reference
Voltage Level
Upper Threshold
Input Signal
Period
T
Triggering occurs when the upper threshold period is reached,
before the rising edge (↑ trigger slope) of the input signal
crosses the reference voltage level.
The trigger point is determined by the upper threshold of the
period range.
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......