171
Detailed Settings for Modules
Refer to “To Open [Each Ch] and Make a Channel Selection” (p. 164).
Level
This is enabled only when
[Mode]
is
[Pulse Width]
or
[Duty].
In pulse width/duty rate measurements, set which level is detected when the
threshold is exceeded.
HIGH
Measures above the threshold value.
(Default setting)
LOW
Measures below the threshold value.
Refer to “To Open [Each Ch] and Make a Channel Selection” (p. 164).
Smoothing
This is enabled only when
[Mode]
is
[Freq]
or
[RPM].
Set smoothing.
Off
The measured data is recorded as is. (A step waveform.)
(Default setting)
On
The measured data is interpolated so that the waveform is smooth
before it is output. (Upper limit: 10 kHz, further delayed from Off.)
Hold
This is enabled only when
[Mode]
is
[Freq]
or
[RPM].
Set hold for frequency and count.
Off
(1Hz/0.5Hz/
0.2Hz/0.1Hz)
When the frequency reaches the value in the brackets but is not settled,
it is defined as a stopped item and the measurement value becomes 0
Hz (0 rpm).
(Default setting)
On
The value defined last time is retained.
8
Utility Functions
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......