241
Display Settings
To get details on each block
The trigger time and measurement status of each block can be viewed on the list.
Move the cursor to
[Map/List]
, and select
[List]
.
Block No.
Blocks can be selected with the
Up/down CURSOR
or
F
keys.
You can move the cursor to the
[Reference Block]
column to set a block as a Reference Block.
To switch block waveforms on the Waveform screen
To select the block you want to see, press the
WAVE
key at the Waveform screen and switch Pos
to Block.
Refer to “7.8 View Block Waveforms” (p. 137).
• When displaying memory division blocks as a list, blocks may have the same trigger times. This
occurs because the minimum resolution of the clock used by this instrument is 1/128th of a second
(7.8125 ms) and measurement occurs during this interval.
Example:
The recording time is 1.25 ms and the displayed
trigger times of blocks 3 and 4 are the same.
•
If triggers occur continuously within an interval shorter than 500 μs, the displayed trigger time may
indicate times slower than the actual triggers.
Memory Division Function
12
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......