8.10.8 Setting Model MR8990 Digital Voltmeter Unit
• When the Model MR8990 Digital Voltmeter Unit is installed as unit 1, 2, 9, or 10, the standard logic
channels cannot be used.
• The resolution of the data measured by the recorder function is 16 bits.
Refer to “To Open [Each Ch] and Make a Channel Selection” (p. 164).
Frequency
Set the power frequency.
Set 50 Hz or 60 Hz according to the power frequency for your region.
50Hz
Cycle 20 ms
(Default setting)
60Hz
Cycle 16.67 ms
If the power frequency setting is not configured correctly, measurement values will be unstable.
NPLC
PLC (Power Line Cycle) is the time equivalent to one cycle of the power frequency.
Set the integration time based on one PLC.
0.1 to 0.9, 1
(
Default setting)
to 10, 20, 30, 40, 50, 60, 70, 80, 90, 100
Example: When the power frequency is 50 Hz, and NPLC is set to 10, 20 ms × 10 =
200 ms.
The measurement data refresh rate becomes 200 ms.
Fast response
Data can be refreshed at high speed.
Off
Refreshes data at the integration time set for NPLC.
(Default setting)
On
Calculates the moving average and refreshes the data at high speed.
Refreshes the data at 0.1 PLC when NPLC is up to 9.
Refreshes the data at 1 PLC when NPLC is 10 or above.
Calibration
This setting is used for performing calibration or synchronization between channels
automatically when measurement is started. Performing synchronization between
channels enables matching the timing for the start of integration.
Off
No calibration and synchronization.
(Default setting)
On
Perform calibration and synchronization.
Synchronization Perform only synchronization between channels.
8
Utility Functions
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......