351
Measurement Functions
18.3.2 Recorder Function
Timebase
10, 20, 50, 100, 200, 500 ms/div
1, 2, 5, 10, 30, 50, 100 s/div
1, 2, 5, 10, 30 min/div, 1 h/div
Time axis resolution
100 points/div
Sampling period
1, 10, 100 µs
1, 10, 100 ms (select from periods that are 1/100 or less of the timebase)
Recording length
•Fixed recording length:
25, 50, 100, 200, 500, 1000, 2000, 5000, 10000, 20000, 50000 div
•Arbitrary recording length:
Can be set in units of divisions (up to 80,000 div)
•Continuous
Screen and printing
settings
1, 2, 4, 8, 16 screens
Waveform storage
The most recent 80,000 divisions of measurement data is retained in internal memory
Waveform scrolling
Left-right scrolling by jog and shuttle knobs.
Backward scrolling during measurement is possible
Real-time printing*
Available with timebase set to 500 ms/div or longer
(Start and stop printing by Function key while recording)
With 10 ms/div to 200 ms/div recording, prints simultaneously with some delay from
recording (Continuous recording length not available)
During continuous recording at 10 ms/div to 200 ms/div, print manually after measurement
stops
Manual print*
By pressing the
key (prints up to 50 mm/s, depending on print coverage)
Selection printing*
Included, Prints the waveform between A and B cursors (Print All and Print Selection)
Report printing*
Included
Auto-save
Automatically saves data on the CF card, built-in drive, USB memory stick after measuring
(binary or text)
When sampling is slow, saving starts during recording.
* When Model U8350 Printer Unit is installed
Specifications
18
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......