369
Module Specifications
Input coupling
AC/DC/GND
Input resistance
1 M
Ω
±1%
Input capacitance
30 pF±10 pF (at 100 kHz)
A/D resolution
12 bits
Maximum sampling
rate
1 MS/s
Input terminals
Insulated BNC terminal
Maximum input
voltage
400 V DC
Maximum rated
voltage to earth
300 V AC, DC (between individual input channels and enclosure as well as between
individual input channels)
Measurement category II (anticipated transient overvoltage: 2500 V)
Operating
temperature and
humidity
As per Memory HiCorder in which Model 8972 is installed
Operating
environment
As per Memory HiCorder in which Model 8972 is installed
Storage temperature
and humidity
-10°C to 50°C (14°F to 122°F), 80% RH or less (no condensation)
Dimensions
Approx. 106W × 19.8H × 196.5D mm (4.17″W × 0.78″H × 7.74″D)
Mass
Approx. 250 g (8.8 oz.)
Effect of radiated
radio-frequency
electromagnetic field
±15% f.s. (max.) at 3 V/m
Effect of conducted
radio-frequency
electromagnetic field
±20% f.s. (max.) at 3 V (100 mV/div range, with 1 V DC input)
Standards
Safety EN61010
EMC
EN61326 Class A
Specifications
18
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......