383
19
Maintenance and Service
WARNING
•
Do not touch the internal parts of this instrument as high voltage is generated
in some parts. Customers are not allowed to modify, disassemble, or repair the
instrument. Doing so may cause fire, electric shock, or injury.
•
Do not touch the internal parts of this instrument as high voltage is generated
in some parts. Customers are not allowed to modify, disassemble, or repair the
instrument. Doing so may cause fire, electric shock, or injury.
The recording paper can be purchased via authorized Hioki distributor or reseller.
Calibrations
The frequency of calibration varies depending on the status of the instrument and installation environment.
We recommend that the frequency of calibration be determined in accordance with the status of the
instrument and installation environment and that you request that calibration be performed periodically.
Precautions During Shipment
Be sure to observe the following precautions:
• To avoid damage to the instrument, remove all accessories and optional equipment from the instrument.
Use the original packing materials in which the product was shipped. Damage that occurs during
transport is not covered by the warranty.
• When sending the instrument for repair, be sure to include details of the problem.
To avoid straining some parts of the printer, and to prevent dirt adhering to the print head, close the printer
cover.
Replaceable Parts and Operating Lifespan
The characteristics of some parts used in the product may deteriorate due to long time of use.
To ensure the product can be used over the long term, it is recommended to replace these parts on a
periodic basis.
Contact your authorized Hioki distributor or reseller for replacement.
The service life of parts varies with the operating environment and frequency of use. Parts are not
guaranteed to operate throughout the recommended replacement cycle.
Part
Recommended
replacement cycle
Remarks/conditions
Fan motor
Approx. 50,000 hours
—
Power supply
Approx. 40,000 hours
—
Print head
After using approx.
1,000 rolls of
recording paper
—
LCD backlight
(Half-life of
brightness)
Approx. 80,000 hours At ambient temperature of 25°C
The service life varies greatly depending on the ambient environment.
The service life becomes shorter under high temperature in particular.
Model U8330 SSD
Unit
Approx. 20,000 hours
(continuous writing)
In ambient environment of 25°C TBW (Total Byte Written): Approx. 60
TB
Data retention period: Approx. one year (When power is OFF)
It is recommended to back up the data at regular intervals.
Electrolytic
capacitors
Approx. 50,000 hours PCBs on which an electrolytic capacitor must be replaced.
19 Maintenance and Service
19
Maintenance and Service
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......