265
Screen Display Settings
13.5.1 Display Running Spectrums
If
[Format]
is set to
[Running spectrum]
, you can observe changes in frequency over time.
Procedure
Press the
STATUS
key to open the Status screen, and then select the
[Status]
sheet.
1
Select the reference data.
Move the cursor to
[Reference]
, and select
[New Data]
.
2
Select the display format.
Move the cursor to
[Format]
, and select
[
Running
spectrum]
.
•
[Running spectrum]
can only be set when
[Reference]
is
[New Data]
.
• The calculation interval (time interval of the running spectrum waveform and the waveform) is not
regulated.
• Averaging cannot be used.
• Calculations that can be analyzed by running spectrum are limited to the following list. When other
calculations are selected or calculation settings are changed after measurement stops, waveforms
will not be displayed.
(Linear spectrum, RMS spectrum, power spectrum, power spectrum density, LPC analysis, transfer
function, cross power spectrum, 1/1 octave analysis, 1/3 octave analysis)
• When running spectrum is displayed, the auto-saving/auto-printing during measurement cannot be
performed.
13
FFT
Function
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......