189
Set the Trigger Mode
9.2 Set the Trigger Mode
This setting controls whether the instrument continues to accept triggers after measuring.
If all trigger sources are disabled (Off, with no trigger setting), measurement starts immediately (free-
running).
Procedure
Press the
DISP
key to open the Waveform screen.
1
Move the cursor to
[Trigger].
2
Select the trigger mode.
Single
Only one trigger is recognized. After pressing the
START
key, once a trigger is applied, a waveform is recorded for the
specified recording length, then measurement stops.
Repeat
Triggers are accepted continuously.
When no trigger is applied, the instrument enters the Trigger
Wait state. Press the
STOP
key to stop measurement. (See
below)
Auto
Triggers are accepted continuously.
If no trigger is applied within about one second, a waveform
of the specified recording length is automatically recorded.
Press the
STOP
key to stop measurement.
Available options depend on the function.
Trigger modes
Function
Single
O
O
(Default setting)
Repeat
O
O
Auto
O
(Default setting)
-
To Stop Measurement, press the
STOP
key.
Press once: Recording will stop at the end of the specified recording length.
Press twice: Recording stops when the key is pressed.
When the trigger mode is set to
[Repeat]
During the processing interval (auto-save, auto-print, waveform display processing, numerical calculation)
from the end of recording until next entering the Trigger Wait state, triggering will not be accepted.
9
Trigger Settings
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......