380
Module Specifications
18.6.13
MR8791 Pulse Generator Unit
General Specifications
Temperature
and humidity for
guaranteed accuracy
23°C±5°C (73°F±9°F), 80% RH or less (no condensation) (When installed in the Memory
HiCorder)
Guaranteed accuracy
period
1 year
Product warranty
period
3 years
Operating temperature
and humidity
As per Memory HiCorder in which Model MR8791 is installed
Operating
environment
As per Memory HiCorder in which Model MR8791 is installed
Storage temperature
and humidity
-20°C to 50°C (-4°F to 122°F), 90% RH or less (no condensation)
Maximum rated
voltage to earth
33 V rms AC or 70 V DC (between output channel and enclosure)
Anticipated transient overvoltage: 330 V
Dimensions
Approx. 106W × 19.8H × 196.5D mm (4.17″W × 0.78″H × 7.74″D) (excluding protrusions)
Mass
Approx. 230 g (8.1 oz.)
Number of output
channels
8 (output channels and enclosure isolated; output units isolated)
(channels not isolated from each other [common GND]) (channels not isolated from output
connector frame [common GND])
Output mode 1
Pattern output/pulse output (switched for all 8 channels at once)
Output mode 2
Logic output:
Output voltage level:
0 V to 5 V (high level 3.8 V or greater, low level 0.8 V or less)
Rated current:
±5 mA
Open collector Output: Collector/emitter absolute maximum rated voltage: 50 V
Over-current protection: 100 mA
Logic output/open-collector output (set separately for each of 8 channels)
Output mode 3
Output/open (= self-test) (switched for all 8 channels at once)
Open-collector
output definition
(rising time [10% to
90%])
5 μs (max.) (load capacity of 1000 pF, pull-up resistance of 1 k
Ω
)
Self-test function
Detected voltages: High level 3.4 V or greater, low level 1.6 V or less
Relay switching time
5 ms or less (logic/open collector switching, output/open [self-test] switching)
Standards
Safety EN61010
EMC
EN61326 Class A
Pulse Output Specifications
Output frequency
Setting range:
0 Hz to 20 kHz (set separately for each of 8 channels)
Setting resolution:
0.1 Hz
Frequency accuracy: ±100 ppm
Duty ratio
Setting range:
0.1% to 99.9%, 0, 100% (DC)
Setting resolution:
0.1%
Duty ratio accuracy: {±100 ppm (setting period) ±150 ns} of setting
Priority to the “Minimum pulse width” and “Open collector output
regulation” specifications
Minimum pulse width
1 μs
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......