191
Analog Signal Triggers
1. Level Trigger
A trigger is applied when the input signal crosses the specified trigger level (threshold voltage).
Trigger Level
Input Waveform
Trigger Slope:
[
↓]
T
T
[
↑]
In this manual,
T
indicates a “trigger point,” the time at which a trigger is applied.
Type
Parameters
[Level]
[L]
(Level)
Sets the level (voltage value) for the trigger. (The setting can be made in 1/50
increments.)
[S]
(Slope)
Determines whether triggering occurs when the signal crosses the threshold (trigger
level) on the rising or falling edge. With the [ ] setting, triggering occurs in both
directions. (↑, ↓, )
[Event]
The number of signal rising or falling edge events is counted and triggering occurs
when the Event number set here is exceeded. (1 to 4000)
[F]
(Filter)
Triggering occurs when the trigger criteria are met within the specified filter width.
This is useful to prevent spurious triggering due to noise. (
: OFF, 0.1
- 10 div
: OFF, ON*) *: Filter width 10 ms.
2. In-Window Trigger, Out-of-Window Trigger
By setting upper and lower limit values for the trigger, a trigger will be applied when the input signal enters
(In) or leaves (Out) this range.
Upper Threshold
Lower Threshold
T
T
(In)
Upper Threshold
Lower Threshold
T
T
(Out)
Type
Parameters
[In]
or
[Out]
[Lower]
Set the lower limit value.
(The setting can be made in 1/50 increments.)
[Upper]
Set the upper limit value.
(The setting can be made in 1/50 increments.)
[F]
(Filter)
Triggering occurs when the trigger criteria are met within the specified filter width.
This is useful to prevent spurious triggering due to noise.
(
: Off, 0.1 - 10 div,
: Off, On*)
*: Filter width 10 ms
9
Trigger Settings
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......